ISO/IEC 17025 Sysmex R-500 hematology reticulocyte analyzer validation

Lab Hematol. 2007;13(2):43-8. doi: 10.1532/LH96.06046.

Abstract

The Sysmex R-500 (R-500) Hematology Analyzer is a bench-top system appropriate for the analysis of limited batches of blood samples. The R-500 provides percentage proportional (RET%), absolute reticulocyte (RET#), and absolute red blood cell (RBC#) counts. The system was validated at the Doping Control Laboratory of Athens, according to the International Committee for Standardization in Hematology, International Standards Organization (ISO/IEC) 17025, and World Antidoping Agency (WADA) specifications. The instrument calibration was performed according to the manufacturer and validation parameters comprised linearity, precision, uncertainty (intermediate and long-term precision), comparability, effect of drift, carryover, stability, and accuracy. The linearity and the comparability studies for RET#, RET%, and RBC# were expressed in regression factors (R2) and coefficients of correlation [r(x, y)], respectively. For the precision studies, the coefficients of variation for RET#, RET%, and RBC# were 9.49%, 9.83%, and <1.5%, respectively. For the intermediate precision studies, the coefficients of variation for RET#, RET%, and RBC# were 3.1%, 3.6%, and 0.6%, respectively. Carryover was found to be negligible. Sample stability was demonstrated at both room temperature and at 4 degrees C over a 24-hour period. Comparability studies for the R-500 were performed using a Sysmex SE-9500. The total evaluation led to the conclusion that the R-500 is an accurate and precise analyzer and because of to its relatively limited size, it can be considered a portable instrument, capable to be used in sports competition and training sites, where doping control and health tests are conducted. The analytical methodology of RET% measurement by the R-500 has been incorporated into the Doping Control Laboratory of Athens' Scope of Accreditation according to the ISO/IEC 17025 and WADA specifications.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Adult
  • Cold Temperature
  • Doping in Sports*
  • Equipment Design
  • Humans
  • Reproducibility of Results
  • Reticulocyte Count / instrumentation*
  • Reticulocyte Count / standards
  • Reticulocytes / ultrastructure
  • Temperature