Depth analysis of polymer-coated steel samples using near-infrared femtosecond laser ablation inductively coupled plasma mass spectrometry

Anal Chem. 2007 Jul 1;79(13):4908-14. doi: 10.1021/ac070241q. Epub 2007 Jun 5.

Abstract

The viability of near-infrared femtosecond laser ablation (fs-LA) inductively coupled plasma mass spectrometry (ICPMS) for the in-depth analysis of polymer coatings over galvanized steel substrates has been studied. A good depth resolution was obtained modifying the femtosecond Gaussian beam to a flat-top beam by using a liquid-crystal display. In order to avoid mixing of information coming from successive shots, a low repetition rate was accomplished and signals were monitored shot by shot. Different kinds of coatings were used to demonstrate the capability of femtosecond ablation for depth-profiling analysis. Ablation was conducted under He atmosphere, after sample cell Ar was admixed. The depth profiles obtained by LA-ICPMS are in good agreement with those obtained by GD-OES for the three analyzed samples. In cases where due to averaging over several millimeter sample roughness determines the depth resolution of GD-OES, it was found that LA-ICPMS achieves better depth resolution due to the better lateral resolution. The depth resolution obtained by LA-ICPMS was found to be 240 nm and 2.3 microm, for a hot-dip galvanized steel (HDGS) and a polymer-polymer-coated HDGS, respectively, compared to the 2.2 and 4.5 microm achieved with GD-OES for the same samples.