Quadrature coil design for high-resolution carotid artery imaging scores better than a dual phased-array coil design with the same volume coverage

J Magn Reson Imaging. 2007 May;25(5):1079-84. doi: 10.1002/jmri.20894.

Abstract

Purpose: To evaluate the ability of a custom-built coil design to provide improved signal-to-noise ratio (SNR) and less signal drop with increasing depth at the carotid artery.

Materials and methods: Phased-array surface coils can provide a high SNR to image the carotid vessel wall. However, given the required field-of-view (FOV) and penetration depth, these coils show either a fast signal drop with increasing depth or a moderate SNR at increased coil size. A quadrature surface coil (a butterfly coil in conjunction with a linear single-loop coil) was compared with a phased-array coil in phantom and human studies.

Results: The phantom studies showed that the quadrature coil has better SNR over the required FOV than a standard phased-array coil (26% at 3 cm depth).

Conclusion: The quadrature coil enables better image quality to be achieved.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Carotid Stenosis / diagnosis*
  • Computer Simulation
  • Equipment Design
  • Humans
  • Magnetic Resonance Angiography / instrumentation*
  • Phantoms, Imaging