Concomitant gradient field effects in balanced steady-state free precession

Magn Reson Med. 2007 Apr;57(4):721-30. doi: 10.1002/mrm.21183.

Abstract

Linear magnetic field gradients spatially encode the image information in MRI. Concomitant gradients are undesired magnetic fields that accompany the desired gradients and occur as an unavoidable consequence of Maxwell's equations. These concomitant gradients result in undesired phase accumulation during MRI scans. Balanced steady-state free precession (bSSFP) is a rapid imaging method that is known to suffer from signal dropout from off-resonance phase accrual. In this work it is shown that concomitant gradient phase accrual can induce signal dropout in bSSFP. The spatial variation of the concomitant phase is explored and shown to be a function of gradient strength, slice orientation, phase-encoding (PE) direction, distance from isocenter, and main field strength. The effect on the imaging signal level was simulated and then verified in phantom and in vivo experiments. The nearest signal-loss artifacts occurred in scans that were offset from isocenter along the z direction with a transverse readout. Methods for eliminating these artifacts, such as applying compensatory frequency or shim offsets, are demonstrated. Concomitant gradient artifacts can occur at 1.5T, particularly in high-resolution scans or with additional main field inhomogeneity. These artifacts will occur closer to isocenter at field strengths below 1.5T because concomitant gradients are inversely proportional to the main field strength.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts*
  • Brain Mapping / methods*
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging