A 96-channel neural stimulation system for driving AIROF microelectrodes

Conf Proc IEEE Eng Med Biol Soc. 2004:2004:4244-7. doi: 10.1109/IEMBS.2004.1404183.

Abstract

We present the design and testing of a 96-channel stimulation system to drive activated iridium oxide (AIROF) microelectrodes within safe charge-injection limits. Our system improves upon the traditional capacitively coupled, symmetric charge-balanced biphasic stimulation waveform so as to maximize charge-injection capacity without endangering the microelectrodes. It can deliver computer-controlled cathodic current pulse for to up to 96 AIROF microelectrodes and positively bias them during the inter-pulse interval. The stimulation system is comprised of (1) 12 custom-designed PCB boards each hosting an 8-channel ASIC chip, (2) a motherboard to communicate between these 12 boards and the PC, (3) the PC interface equipped with a DIO card and the corresponding software. We plan to use this system in animal experiments for intracortical neural stimulation of implanted electrodes within our visual prosthesis project.