Evaluation of the performance of the automated NucliSENS easyMAG and EasyQ systems versus the Roche AmpliPrep-AMPLICOR combination for high-throughput monitoring of human immunodeficiency virus load

J Clin Microbiol. 2007 Apr;45(4):1244-9. doi: 10.1128/JCM.01540-06. Epub 2007 Jan 31.

Abstract

This study presents the data of an evaluation of the automated Nuclisens easyMAG and EasyQ systems versus the Roche AmpliPrep-AMPLICOR combination for testing of high-volume human immunodeficiency virus (HIV) load. This represents a follow-up of a previous study investigating the performance of the real-time Nuclisens assay using the semiautomated NucliSENS miniMAG extraction procedure. Three hundred eighteen patient samples were analyzed using both methods. The easyMAG-EasyQ HIV type 1 system has a higher sensitivity and broader dynamic range than the Cobas AmpliPrep-AMPLICOR system when the standard Roche assay is used alone, 25 to 3,000,000 IU/ml versus 400 to 750,000 HIV RNA copies/ml, respectively. There was significant correlation between the assays (0.93; P < 0.0001), with good accuracy (percent similarity mean mu = 96%), good precision (percent similarity standard deviation = 4.97%), and overall good agreement with a low percent similarity coefficient of variation of 5.17 to 6.11%. Bland-Altman analysis revealed that the AMPLICOR assay generated higher values than the EasyQ combination, with 95% of results within clinically acceptable limits. The throughput of samples was greatly improved using the easyMAG-EasyQ system, allowing 144 samples to be processed within 6 h. The potential for contamination has been dramatically reduced using the automated extraction system. Additional negative controls have been added to the kit to monitor for contamination based on the South African experience. This assay thus presents a real option for monitoring HIV load assays in high-volume testing environments.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • HIV Infections / virology*
  • HIV-1 / growth & development*
  • Humans
  • RNA, Viral / blood
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Viral Load / methods*

Substances

  • RNA, Viral