Ion-trap tandem mass spectrometry-based analytical methodology for the determination of polychlorinated biphenyls in fish and shellfish. Performance comparison against electron-capture detection and high-resolution mass spectrometry detection

J Chromatogr A. 2007 Feb 23;1142(2):199-208. doi: 10.1016/j.chroma.2006.12.068. Epub 2006 Dec 23.

Abstract

Optimization of the Varian Saturn 2200 ion-trap tandem mass spectrometry (IT-MS/MS) system and comparison of its data quality with two other detection methods [electron-capture detection (ECD) and high-resolution mass spectrometry (HRMS)] was pursued by measuring polychlorinated biphenyls (PCBs) levels in fish and shellfish samples. IT-MS/MS methodology provided limits of detection (LOD) comparable to those obtained by ECD but superior specificity for the detection of a selected number of 39 PCB native congeners and 9 (13)C-labelled PCB standards. The method detection limits (MDLs) established for IT-MS/MS ranged between 1.0 and 5.0 pg/g on a wet weight basis while those obtained by ECD and HRMS were 1.0-4.0 pg/g and 0.1-2.0 pg/g, respectively. Overall, the results obtained in the study demonstrate that gas chromatography (GC) combined with IT-MS/MS provide higher data quality than those achievable by GC-ECD. For this particular set of target analytes the specificity achievable with IT-MS/MS was comparable to that obtained by HRMS and both techniques provided comparable data in terms of accuracy and precision.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Fishes*
  • Polychlorinated Biphenyls / analysis*
  • Quality Control
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Shellfish / analysis*
  • Tandem Mass Spectrometry / methods*

Substances

  • Polychlorinated Biphenyls