Functionalization of PVC membrane with ss oligonucleotides for a potentiometric biosensor

Biosens Bioelectron. 2007 May 15;22(11):2712-7. doi: 10.1016/j.bios.2006.11.014. Epub 2006 Dec 22.

Abstract

A novel application of a single stranded (ss) oligonucleotide as an active component of polymeric membrane in an ion-selective electrode (ISE) is described. The original oligonucleotides, oligo(dA)(15), modified by cholesterol, triphenylmethyl and hexadecyl derivatives, were immobilized into poly(vinyl chloride) (PVC) membrane using extraction protocol. In parallel, the adsorption protocol was used to immobilize unmodified oligo(dA)(15) on the PVC membrane based on tridodecylmethyammonium chloride (TDDMA(+)Cl(-)). Immobilization of ss oligonucleotide probe through spacer was more effective for the potentiometric detection of the hybridization between complementary oligonucleotides. It was found that cholesterol-oligo(dA)(15) modified membranes were sensitive toward complementary oligo(dT)(15) in the concentration range 2-80 nM at pH 7. An explanation for the detection mechanism is proposed.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Biosensing Techniques / methods
  • Coated Materials, Biocompatible / chemistry
  • DNA Probes / chemistry*
  • DNA Probes / genetics
  • Electrochemistry / instrumentation*
  • Electrochemistry / methods
  • Equipment Design
  • Equipment Failure Analysis
  • In Situ Hybridization / instrumentation*
  • In Situ Hybridization / methods
  • Oligonucleotides / analysis*
  • Oligonucleotides / chemistry*
  • Oligonucleotides / genetics
  • Polyvinyl Chloride / chemistry*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Coated Materials, Biocompatible
  • DNA Probes
  • Oligonucleotides