Diagnostic accuracy of high-resolution MRI using a microscopy coil for patients with presumed DCIS following mammography screening

J Magn Reson Imaging. 2007 Jan;25(1):96-103. doi: 10.1002/jmri.20809.

Abstract

Purpose: To prospectively evaluate the accuracy of high-resolution (HR)-MRI as a secondary examination in women with abnormal calcifications detected on mammography.

Materials and methods: We used a 4.7-cm microscopy coil to acquire HR-MRI signal data. We examined 52 women with breast lesions preoperatively using HR-MRI and vacuum-assisted core needle biopsy. The lesions were suspicious of malignancy, classified as category 3-5 on mammography (Breast Imaging Reporting and Data System [BI-RADS]), and without a palpable mass. All visualized suspicious lesions were correlated with histological findings. We compared the HR-MRI and pathological findings and calculated the sensitivity, specificity, and accuracy.

Results: We compared the breast HR-MRI results with the gold standard of pathological results for studies of malignancy (DCIS and invasive cancer), and found a sensitivity of 88.5%, specificity of 92.3%, and accuracy of 90.4%. The positive predictive value (PPV) was 92%, and the negative predictive value (NPV) was 88.9%. When breast MRI was compared with pathological results for studies that diagnosed DCIS only, the results revealed a sensitivity of 88.6%, specificity of 88.2%, accuracy of 88.5%, PPV of 93.9%, and NPV of 78.9%.

Conclusion: HR-MRI using a microscopy coil is a useful, reliable, safe, and minimally invasive procedure that is a good choice for secondary assessment of abnormal calcification in the breast.

MeSH terms

  • Adult
  • Aged
  • Biopsy, Needle
  • Breast Neoplasms / diagnosis*
  • Breast Neoplasms / diagnostic imaging
  • Breast Neoplasms / pathology
  • Carcinoma, Intraductal, Noninfiltrating / diagnosis*
  • Carcinoma, Intraductal, Noninfiltrating / diagnostic imaging
  • Carcinoma, Intraductal, Noninfiltrating / pathology
  • Contrast Media
  • Diagnosis, Differential
  • Female
  • Gadolinium DTPA
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Mammography
  • Middle Aged
  • Predictive Value of Tests
  • Prospective Studies
  • Sensitivity and Specificity

Substances

  • Contrast Media
  • gadodiamide
  • Gadolinium DTPA