Determination of captopril in human plasma, using solid phase extraction and high-performance liquid chromatography, coupled to mass spectrometry: application to bioequivalence study

J Chromatogr B Analyt Technol Biomed Life Sci. 2007 May 1;850(1-2):59-67. doi: 10.1016/j.jchromb.2006.11.007. Epub 2006 Dec 4.

Abstract

A specific high performance liquid chromatography-mass spectrometric (LC-MS/MS) assay was developed for the determination of captopryl in plasma. The retention time was 1.45 and 1.37 min for captopril and enalapril, respectively. The overall mean recovery, using SPE extraction with OASIS HLB cartridges, was found to be 107.2+/-9.5 and 100.04+/-2%, respectively. Calibration curves were linear in the concentration range of 10.00-2000.00 ng/ml, and the lower limit of quantification (LLOQ) was 10.00 ng/ml. The LLOQ was sensitive enough for detecting terminal phase concentrations of the drug. Inter-batch precision of the method ranged from 0.88 to 1.95%. Intra-batch accuracy ranged from 97.15 to 105.77%, while intra-batch precision ranged from 2.49 to 5.66% at concentrations of 30.00, 760.00 and 1500.00 ng/ml. The developed method was applied to study bioequivalence of captopril in a group of 25 human subjects at a single oral dose of a 50mg tablet.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Angiotensin-Converting Enzyme Inhibitors / blood*
  • Angiotensin-Converting Enzyme Inhibitors / pharmacokinetics
  • Calibration
  • Captopril / blood*
  • Captopril / pharmacokinetics
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Tandem Mass Spectrometry / methods*
  • Therapeutic Equivalency

Substances

  • Angiotensin-Converting Enzyme Inhibitors
  • Captopril