Prostate biopsies guided by three-dimensional real-time (4-D) transrectal ultrasonography on a phantom: comparative study versus two-dimensional transrectal ultrasound-guided biopsies

Eur Urol. 2007 Oct;52(4):1097-104. doi: 10.1016/j.eururo.2006.11.034. Epub 2006 Nov 27.

Abstract

Objective: This study evaluated the accuracy in localisation and distribution of real-time three-dimensional (4-D) ultrasound-guided biopsies on a prostate phantom.

Methods: A prostate phantom was created. A three-dimensional real-time ultrasound system with a 5.9MHz probe was used, making it possible to see several reconstructed orthogonal viewing planes in real time. Fourteen operators performed biopsies first under 2-D then 4-D transurethral ultrasound (TRUS) guidance (336 biopsies). The biopsy path was modelled using segmentation in a 3-D ultrasonographic volume. Special software was used to visualise the biopsy paths in a reference prostate and assess the sampled area. A comparative study was performed to examine the accuracy of the entry points and target of the needle. Distribution was assessed by measuring the volume sampled and a redundancy ratio of the sampled prostate.

Results: A significant increase in accuracy in hitting the target zone was identified using 4-D ultrasonography as compared to 2-D. There was no increase in the sampled volume or improvement in the biopsy distribution with 4-D ultrasonography as compared to 2-D.

Conclusion: The 4-D TRUS guidance appears to show, on a synthetic model, an improvement in location accuracy and in the ability to reproduce a protocol. The biopsy distribution does not seem improved.