Studies on PVP hydrogel-supported luminol chemiluminescence: 2. Luminometer calibration and potential analytical applications

Luminescence. 2007 Mar-Apr;22(2):126-33. doi: 10.1002/bio.935.

Abstract

The use of poly(N-vinyl-2-pyrrolidone) (PVP) hydrogel-supported luminol chemiluminescence (CL) for the automatic determination of hydrogen peroxide and the quantification of the antiradical capacity of Trolox is described. The hydrogel containing luminol and hemin is prepared directly on a 96-well microplate and can be stored for up to 3 months without significant decrease in CL quantum yields. Furthermore, this system can also be used as a secondary light standard for the calibration of microplate luminometers.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antioxidants / chemistry
  • Antioxidants / pharmacology
  • Automation
  • Calibration
  • Chromans / chemistry
  • Chromans / pharmacology
  • Hemin / chemistry
  • Hydrogels / chemistry*
  • Hydrogen Peroxide / analysis*
  • Luminescent Measurements / instrumentation*
  • Luminescent Measurements / methods
  • Luminol / chemistry*
  • Molecular Structure
  • Povidone / chemistry*

Substances

  • Antioxidants
  • Chromans
  • Hydrogels
  • Luminol
  • Hemin
  • Hydrogen Peroxide
  • Povidone
  • 6-hydroxy-2,5,7,8-tetramethylchroman-2-carboxylic acid