A simple and rapid determination of biapenem in plasma by high-performance liquid chromatography

J Chromatogr B Analyt Technol Biomed Life Sci. 2006 Nov 21;844(1):148-52. doi: 10.1016/j.jchromb.2006.06.023. Epub 2006 Jul 27.

Abstract

A simple, rapid and precise HPLC method using ultrafiltration to remove plasma protein was developed to determine biapenem concentrations in human plasma. Plasma was separated by centrifugation at 4 degrees C from blood collected in heparinized vacuum tubes, and biapenem was stabilized by immediate mixing the plasma with 1M 3-morpholinopropanesulfonic acid (MOPS) buffer (pH 7.0) (1:1). Biapenem was detected by ultraviolet absorbance at 300nm with no interfering plasma peak. The calibration curve of biapenem in human plasma was linear from 0.04 to 50microg/mL. The limit of detection was 0.01microg/mL, which was more than 40-fold lower than that of conventional plasma protein precipitation using ammonium sulfate. The assay has been clinically applied to pharmacokinetic studies in patients.

Publication types

  • Validation Study

MeSH terms

  • Anti-Bacterial Agents / blood*
  • Anti-Bacterial Agents / pharmacokinetics
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Reference Standards
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet
  • Thienamycins / blood*
  • Thienamycins / pharmacokinetics

Substances

  • Anti-Bacterial Agents
  • Thienamycins
  • biapenem