Estimation of base blood pressure by using a new device in the outpatient clinic

Hypertens Res. 2006 Apr;29(4):233-41. doi: 10.1291/hypres.29.233.

Abstract

Direct measurement of intra-arterial blood pressure (BP) for 24-h provides approximately 100,000 values that vary enormously, but each (BPi) can be expressed by the equation BPi = BP0 + DeltaBPi (BP0, base BP; DeltaBPi, BP increment, i=1, 2, ..., 100 x 10(3)). About 20% of outpatients with hypertension exhibit white-coat hypertension (WCH). In such patients, DeltaBPc (i = c; c, time at the clinic) is surmised to be large. A method for explaining the physiological factors in DeltaBPc and the estimation of base BP in the outpatient clinic is important. This study addresses this issue. A total of 293 subjects were divided into four groups: 1) WCH group, 45 individuals (office BP > or = 140/90 mmHg and 24-h indirect BP < 125/80 mmHg); 2) normotensive (NT) group, 84 controls matched for age and sex; 3) WHO-I group, 95 hypertensive patients with WHO stage I (office BP > or = 140/90 mmHg and 24-h BP > or = 125/80 mmHg); and 4) WHO-II group, 69 hypertensive patients with WHO stage II. Their BPc and heart rate (HR; HRc, clinic HR) values were measured by a BP-ECG monitoring device in the outpatient clinic. Power-spectral analysis was used to obtain the ratio between the low-frequency component (LF) and high-frequency component (HF) of ECG-RR variability (LF/HF = LH). Twenty-four-hour indirect BP (and BP0) and base HR (HR0) were measured by a portable device (TM2425) at 30-min intervals. Then, DeltaBPc (= BPc - BP0) was estimated by performing linear multivariate analysis applying the model equation DeltaBPc = (BPc -alphaLH)(1-betaHR0/HRc) + epsilon to the above variables (alpha and beta, constant values; epsilon, error). This model equation made it possible to estimate BP0 (and DeltaBPc) with a high coefficient of correlation (r > or = 0.85, mean of error less than 0.82 +/- 5.9 mmHg). The predictive accuracy for discrimination between WCH and sustained hypertension (WHO-I and WHO-II groups) by this equation was 88%. The new DeltaBP-estimation device (BP-ECG monitor) enabled us to infer BP0 and is therefore useful in estimating WCH in the outpatient clinic.

Publication types

  • Controlled Clinical Trial

MeSH terms

  • Adult
  • Aged
  • Ambulatory Care Facilities
  • Baroreflex
  • Blood Pressure Monitoring, Ambulatory / instrumentation*
  • Blood Pressure Monitoring, Ambulatory / methods
  • Electrocardiography / instrumentation*
  • Electrocardiography / methods
  • Female
  • Heart Rate
  • Humans
  • Hypertension / diagnosis*
  • Linear Models
  • Male
  • Middle Aged
  • Models, Cardiovascular
  • Multivariate Analysis