Goldvalve detachable balloon: "in vitro" assessment of safety and imaging artifacts in a 3-T MR system

Neuroradiology. 2006 Jun;48(6):409-11. doi: 10.1007/s00234-006-0078-z. Epub 2006 Apr 26.

Abstract

The Goldvalve balloon is the only currently available detachable balloon. We undertook a study to determine safety and imaging artifacts in a MR environment. We found no evidence for heating or deflection of the balloon in a comprehensive series of in vitro experiments at 3 T. MR imaging at field strengths up to 3 T of patients with implanted Goldvalve balloons is safe. Imaging artifacts are minimal.

MeSH terms

  • Artifacts*
  • Catheterization / instrumentation*
  • Equipment Design
  • Equipment Safety
  • Hot Temperature
  • Humans
  • Magnetic Resonance Imaging*
  • Models, Cardiovascular