Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography

Ophthalmology. 2006 Jun;113(6):937-41. doi: 10.1016/j.ophtha.2006.01.063.

Abstract

Purpose: To compare central corneal thickness measurements and their reproducibility when taken by a rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography/pachymetry.

Design: Experimental study.

Participants: Seventy-four eyes of 64 subjects without ocular abnormalities other than cataract.

Methods: Corneal thickness measurements were compared among the 3 methods in 54 eyes of 54 subjects. Two sets of measurements were repeated by a single examiner for each pachymetry in another 10 eyes of 5 subjects, and the intraexaminer repeatability was assessed as the absolute difference of the first and second measurements. Two experienced examiners took one measurement for each pachymetry in another 10 eyes of 5 subjects, and the interexaminer reproducibility was assessed as the absolute difference of the 2 measurements of the first and second examiners.

Main outcome measures: Central corneal thickness measurements by the 3 methods, absolute difference of the first and second measurements by a single examiner, absolute difference of the 2 measurements by 2 examiners, and relative amount of variation.

Results: The average measurements of central corneal thickness by a rotating Scheimpflug camera, scanning-slit topography, and ultrasonic pachymetry were 538+/-31.3 microm, 541+/-40.7 microm, and 545+/-31.3 microm, respectively. There were no statistically significant differences in the measurement results among the 3 methods (P = 0.569, repeated-measures analysis of variance). There was a significant linear correlation between the rotating Scheimpflug camera and ultrasonic pachymetry (r = 0.908, P<0.0001), rotating Scheimpflug camera and scanning-slit topography (r = 0.930, P<0.0001), and ultrasonic pachymetry and scanning-slit topography (r = 0.887, P<0.0001). Ultrasonic pachymetry had the smallest intraexaminer variability, and scanning-slit topography had the largest intraexaminer variability among the 3 methods. There were similar variations in interexaminer reproducibility among the 3 methods.

Conclusions: Mean corneal thicknesses were comparable among rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit topography with the acoustic equivalent correction factor. The measurements of the 3 instruments had significant linear correlations with one another, and all methods had highly satisfactory measurement repeatability.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Body Weights and Measures
  • Cataract / complications
  • Cornea / anatomy & histology*
  • Cornea / diagnostic imaging
  • Corneal Topography
  • Diagnostic Techniques, Ophthalmological / standards*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Photography
  • Prospective Studies
  • Reproducibility of Results
  • Ultrasonography