Fracture of Colvin-Galloway Future band causing a tear in the anterior mitral leaflet

Ann Thorac Surg. 2006 May;81(5):1879-80. doi: 10.1016/j.athoracsur.2005.05.093.

Abstract

The first case of a fracture of a semi-rigid annuloplasty device is reported. A 56-year-old man presented with recurrence of severe mitral and tricuspid regurgitation 10 months after proper implantation of a Colvin-Galloway Future band (Medtronic Inc, Minneapolis, MN). During redo surgery we made the unexpected finding that its metal skeleton was fractured near the posterior segment. A tear in the anterior mitral leaflet was also found. It is speculated that the device fracture may have put additional strain on the native annulus.