Comparison of noncontact and fiber-based fluorescence-mediated tomography

Opt Lett. 2006 Mar 15;31(6):769-71. doi: 10.1364/ol.31.000769.

Abstract

We present a comparative experimental phantom study of fiber-based and noncontact fluorescence tomography with respect to quantitation and localization of reconstructed fluorescent inclusions in turbid media such as tissue. Noncontact acquisition is usually considered potentially superior to fiber-based techniques because of the availability of a large number of detector readouts through a CCD. Our results indicate, however, that noncontact acquisition itself might improve the quality of reconstructions significantly, even without increasing the number of detectors and thus keeping the inverse problem moderately complex.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Fiber Optic Technology / instrumentation*
  • Image Interpretation, Computer-Assisted / methods*
  • Microscopy, Fluorescence / instrumentation*
  • Microscopy, Fluorescence / methods*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, Optical / instrumentation*