Solid contact potentiometric sensors for trace level measurements

Anal Chem. 2006 Feb 15;78(4):1318-22. doi: 10.1021/ac050749y.

Abstract

A simple procedure for the development of a range of polymeric ion-selective electrodes (ISEs) with low detection limits is presented. The electrodes were prepared by using a plasticizer-free methyl methacrylate-decyl methacrylate copolymer as membrane matrix and poly(3-octylthiophene) as intermediate layer deposited by solvent casting on gold sputtered copper electrodes as a solid inner contact. Five different electrodes were developed for Ag+, Pb2+, Ca2+, K+, and I-, with detection limits mostly in the nanomolar range. In this work, the lowest detection limits reported thus far with solid contact ISEs for the detection of silver (2.0 x 10(-9) M), potassium (10(-7) M), and iodide (10(-8) M) are presented. The developed electrodes exhibited a good response time and excellent reproducibility.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Ion-Selective Electrodes
  • Potentiometry / methods*
  • Trace Elements / analysis*

Substances

  • Trace Elements