Performance of the Colson MAM BP 3AA1-2 automatic blood pressure monitor according to the European Society of Hypertension validation protocol

Rev Port Cardiol. 2005 Nov;24(11):1341-51.
[Article in English, Portuguese]

Abstract

Objective: To evaluate the performance of the Colson MAM BP 3AA1-2 oscillometric automatic blood pressure monitor according to the validation protocol of the European Society of Hypertension, testing its suitability for self-measurement of blood pressure. The performance of the device was assessed in relation to various clinical variables, including age, gender, body mass index, arm circumference and arterial stiffness.

Method: 33 subjects (15 men and 18 women), with a mean age of 47 +/- 10 years, were studied according to the procedures laid down in the European Society of Hypertension validation protocol. Sequential same-arm blood pressure measurements were made, alternating between a mercury standard and the automatic device. The differences among the test-control measurements were assessed and divided into categorization zones of 5, 10 and 15 mmHg discrepancy. Aortic pulse wave velocity was assessed in all subjects with a Complior device (Colson, Paris).

Results: The Colson MAM BP 3AA1-2 passed all three phases of the protocol for both systolic and diastolic blood pressure. The mean differences between the test and control measurements were -1.0 +/- 5.0 mmHg for systolic blood pressure and -1.1 +/- 4.1 mmHg for diastolic blood pressure. Both standard deviations are well below the 8 mmHg limit proposed by the Association for the Advancement of Medical Instrumentation. The predictive value of various clinical variables for the discrepancies was assessed by a regression model analysis, with no variable being found that independently undermined the performance of the monitor. In another regression analysis, we found a similar relation between test and control blood pressures and aortic pulse wave velocity, a widely recognized and validated index of target organ damage.

Conclusions: These data show that the Colson MAM BP 3AA1-2 satisfies the quality requirements proposed by the European Society of Hypertension, demonstrating its suitability for inclusion in integrated programs of clinical surveillance based on self-measurement of blood pressure. The uniformity of its performance over a wide spectrum of clinical characteristics and the relation found with pulse wave velocity further reinforce its clinical validity.