A large ultrasonic bounded acoustic pulse transducer for acoustic transmission goniometry: modeling and calibration

J Acoust Soc Am. 2006 Jan;119(1):54-64. doi: 10.1121/1.2133683.

Abstract

A large, flat ultrasonic transmitter and a small receiver are developed for studies of material properties in acoustic transmission goniometry. While the character of the wave field produced by the transmitter can be considered as a plane wave as observed by the receiver, diffraction effects are noticeable near critical angles and result in the appearance of weak but detectable arrivals. Transmitted ultrasonic waveforms are acquired in one elastic silicate glass and two visco-elastic acrylic glass sample plates as a function of the angle of incidence. Phase velocities are determined from modeling of the shape of curves of the observed arrival times versus angle of incidence. The waveform observations are modeled using a phase propagation technique that incorporates full wave behavior including attenuation. Subtle diffraction effects are captured in addition to the main bounded pulse propagation. The full propagation modeling allows for various arrivals to be unambiguously interpreted. The results of the plane wave solution are close to the full wave propagation modeling without any corrections to the observed wave field. This is an advantage as it places confidence that later analyses can use simpler plane wave solutions without the need for additional diffraction corrections. A further advantage is that the uniform bounded acoustic pulse allows for the detection of weak arrivals such as a low energy edge diffraction observed in our experiments.