A prototype manipulator for magnetic resonance-guided interventions inside standard cylindrical magnetic resonance imaging scanners

J Biomech Eng. 2005 Nov;127(6):972-80. doi: 10.1115/1.2049339.

Abstract

The aim of this work is to develop a remotely controlled manipulator to perform minimally invasive diagnostic and therapeutic interventions in the abdominal and thoracic cavities, with real-time magnetic resonance imaging (MRI) guidance inside clinical cylindrical MR scanners. The manipulator is composed of a three degree of freedom Cartesian motion system, which resides outside the gantry of the scanner, and serves as the holder and global positioner of a three degree of freedom arm which extends inside the gantry of the scanner At its distal end, the arm's end-effector can carry an interventional tool such as a biopsy needle, which can be advanced to a desired depth by means of a seventh degree of freedom. These seven degrees of freedom, provided by the entire assembly, offer extended manipulability to the device and a wide envelope of operation to the user, who can select a trajectory suitable for the procedure. The device is constructed of nonmagnetic and nonconductive fiberglass, and carbon fiber composite materials, to minimize artifacts and distortion on the MR images as well as eliminate effects on its operation from the high magnetic field and the fast switching magnetic field gradients used in MR imaging. A user interface was developed for man-in-the-loop control of the device using real-time MR images. The user interface fuses all sensor signals (MR and manipulator information) in a visualization, planning, and control command environment. Path planning is performed with graphical tools for setting the trajectory of insertion of the interventional tool using multislice and/or three dimensional MR images which are refreshed in real time. The device control is performed with an embedded computer which runs real-time control software. The manipulator compatibility with the MR environment and image-guided operation was tested on a 1.5 T MR scanner.

Publication types

  • Evaluation Study
  • Research Support, N.I.H., Extramural

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Humans
  • Image Interpretation, Computer-Assisted / instrumentation*
  • Image Interpretation, Computer-Assisted / methods
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods
  • Pilot Projects
  • Robotics / instrumentation*
  • Robotics / methods
  • Surgery, Computer-Assisted / instrumentation*
  • Surgery, Computer-Assisted / methods