Nitinol Carbofilm coated stents for peripheral applications: study in the porcine model

J Mater Sci Mater Med. 2005 Dec;16(12):1231-8. doi: 10.1007/s10856-005-4733-4.

Abstract

Purpose: Testing the safety and foreign body reaction (FBR) of a nitinol self-expandable carbon-coated stent system in the porcine animal model at different follow-up (FU) periods.

Methods: Fifteen minipigs received 30 carbon-coated self-expandable nitinol stents in iliac arteries. Explants were carried out at 7 (3 animals), 30 (4 animals), 90 (4 animals) and 180 (4 animals) day FU, for evaluation of acute, sub-acute and chronic biological response to the implanted devices. Histological, immunohistochemical, histomorphometric and scanning electron microscopy (SEM) analyses were performed to assess inflammatory reaction, endothelialization process, neointimal growth and cellular composition.

Results: Thirty stents were successfully implanted. No mural thrombi were observed at gross examination or by angiography. Histologically no significant inflammatory reaction was detected: the stents appeared covered by a thin monolayer of endothelial cells even at 7 day FU. The neointima presented homogeneous growth and moderate thickness after 30, 90 and 180 days explants (0.38+/- 0.36 mm, 0.33+/- 0.30 mm, 0.27+/- 0.25 mm respectively). Internal and external elastic laminae were intact in 95% of stented arteries. Histological data validations of vessel endothelialization was obtained with SEM for the seven day follow-up group.

Conclusions: This study showed good remarkable technical performances, minimal FBR and biocompatibility comparable with other available pre-clinical experimentations.

MeSH terms

  • Alloys*
  • Animals
  • Foreign-Body Reaction*
  • Immunohistochemistry
  • Microscopy, Electron, Scanning
  • Models, Animal*
  • Stents*
  • Swine
  • Swine, Miniature