Study of hip joint dislocation after total hip arthroplasty

Int Orthop. 2006 Feb;30(1):26-30. doi: 10.1007/s00264-005-0032-4. Epub 2005 Dec 13.

Abstract

The present study was undertaken to identify the factors responsible for hip joint dislocation after total hip arthroplasty, laying emphasis on analysis of the background variables of the patients. Of the 317 hips included in the study, ten (3.2%) dislocated. Only the anteversion angle of the cup differed significantly between the dislocation group and the dislocation-free group. The safe zone of the anteversion angle seems to be between 20 and 30 degrees. but it is also essential to set the antetorsion angle of the stem to match the shape of individual bones to create a more stable hip joint. This safe zone may be expanded by the additive effect of antetorsion angle of the stem.

Le but de cette étude a été d’identifier les facteurs responsables d’uneluxation de la hanche après arthroplastie totale en insistant sur l’analysedes différents paramètres de chaque patient. Sur les 317 hanches inclusesdans cette étude, 10 (3,2%) ont présenté une luxation. La seule différencesignificative entre le groupe des sujets ayant présenté une luxation et ceuxn’en ayant pas présenté est l’angle d’antéversion de la cupule. L’angleoptimum d’antéversion de la cupule semble être compris entre 20 et 30 degrésmais il est également essentiel de lier l’angle d’antéversion avec l’angled’antétorsion de la pièce fémorale, celles-ci ayant été manufacturées surmesure et adaptées à la géométrie osseuse de chaque individu de façon àcréer une hanche parfaitement stable. Cette zone de sécurité peut êtreaméliorée par l’addition, à l’angle d’antéversion de la cupule de l’angled’antétorsion de la pièce fémorale.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip / adverse effects*
  • Female
  • Hip Dislocation / etiology*
  • Hip Prosthesis
  • Humans
  • Male
  • Prosthesis Design