Comparison of manual and automated ELISA methods for serum ferritin analysis

J Clin Lab Anal. 2005;19(5):196-8. doi: 10.1002/jcla.20077.

Abstract

Serum ferritin concentration is a sensitive measure of body iron stores. The aim of this study was to compare the performance of two commercially available enzyme-linked immunoassays (ELISAs) for serum ferritin: a widely used manual assay kit (Spectro Ferritin MT), and a new fully automated assay (Immulite). We analyzed serum samples from Moroccan school-aged children (n=51) from a rural area with a high prevalence of iron deficiency anemia (IDA). Four replicates of each sample were analyzed using both assays. For the manual method, the interassay repeatability was 24%, 22%, and 11%, and intraassay precision was 18.3%, 9.2%, and 9.1% at increasing serum ferritin concentrations. Using the automated assay, the interassay repeatability was 7%, 6%, and 6%, and intraassay precision was 1.5%, 5.4%, and 5.5% at increasing serum ferritin concentrations. The two assays were well correlated (y=1.16x+1.83; r=0.98). However, the limits of agreement (LOAs) were wide, particularly at low concentrations. A comparison of the assay results with recommended cutoffs for serum ferritin generated sharply different estimates of the prevalence of iron deficiency (ID) in the sample. We conclude that the automated assay has several potential advantages compared to the manual method, including better precision, less operator dependence, and faster sample through-put.