Measurement of dose-area product with GafChromic XR Type R film

J Appl Clin Med Phys. 2005 Summer;6(3):122-32. doi: 10.1120/jacmp.v6i3.2047. Epub 2005 Aug 12.

Abstract

Many of the newer X-ray machines are equipped with electronic means to provide dose-area product (DAP) information. For machines without that ability, an alternative method is to record radiation on a film that can handle a large amount of cumulative exposure. The use of GafChromic XR Type R film was investigated for this purpose by placing it at the X-ray tube assembly to record the radiation in interventional radiological procedures. Dose-area product was determined with a reflective densitometer and then with a flatbed scanner. Precisions were demonstrated to be 5% and 2%, respectively. In a comparison with the machine-recorded DAP, a regression analysis showed the validity of both techniques for values less than 1200 Gy-cm2.

Publication types

  • Evaluation Study

MeSH terms

  • Algorithms*
  • Densitometry / methods*
  • Dose-Response Relationship, Radiation
  • Equipment Failure Analysis
  • Film Dosimetry / instrumentation*
  • Film Dosimetry / methods*
  • Radiation Dosage
  • Reproducibility of Results
  • Sensitivity and Specificity