Purpose: To compare the signal characteristics and bolus dynamics of 1.0 M gadobutrol and 0.5 M Gd-DTPA for time-resolved, three-dimensional, contrast-enhanced (CE) MRA of the upper torso.
Materials and methods: Ten healthy volunteers were examined with time-resolved three-dimensional CE-MRA (scan time per three-dimensional data set: 0.86 second; voxel size: 3.6 x 2 x 6.3 mm(3)). Each volunteer underwent eight individual examinations after intravenous injection of 0.05 and 0.1 mmol/kg body weight (b.w.) of 1.0 M gadobutrol and 0.5 M Gd-DTPA using two injection rates (2.5 and 5 mL/second). The data analysis included quantitative measurements of the peak signal-to-noise ratio (SNR) and bolus dispersion (full width at half maximum (FWHM)) in the pulmonary artery, left atrium, and thoracic and abdominal aortas.
Results: No significant differences in the peak SNR and bolus dispersion were observed between gadobutrol and Gd-DTPA for all dose levels and injection rates in any of the vascular segments. For both contrast agents a dose of 0.1 mmol/kg b.w. injected with 5 mL/second achieved the highest SNR in all vascular segments.
Conclusion: For the imaging parameters used in this study, higher-concentrated gadolinium chelates offer no relevant advantages for time-resolved three-dimensional CE-MRA of the upper torso.