Visual acuity measurement of prosthetic vision: a virtual-reality simulation study

J Neural Eng. 2005 Mar;2(1):S135-45. doi: 10.1088/1741-2560/2/1/015. Epub 2005 Feb 22.

Abstract

A virtual-reality simulation tested prosthetic visual acuity for both rectangular and hexagonal phosphene grids. Thirteen normally sighted, untrained subjects were required to identify the Landolt C optotype over ten sessions. Overall performance, performance by filter setting (image analysis), and performance by size and orientation of the Landolt C optotype are reported. The results indicated that the hexagonal grid had a slight (4.1%) but statistically significant (p < 0.0001) performance advantage over the rectangular grid for correct identification of the testing symbol. It was also observed that both the phosphene grids had distinct performance profiles relating to their respective spatial orientation. Over the course of the ten sessions, learning afforded subjects an averaged improved performance of 10%.

Publication types

  • Evaluation Study

MeSH terms

  • Artificial Intelligence*
  • Computer Simulation
  • Computer Systems
  • Data Display
  • Environment
  • Equipment Failure Analysis / methods*
  • Humans
  • Online Systems
  • Sensory Aids*
  • Treatment Outcome
  • User-Computer Interface*
  • Vision Disorders / diagnosis*
  • Vision Disorders / rehabilitation*
  • Vision Tests / methods*
  • Visual Acuity / physiology*