Ghosting caused by bulk charge trapping in direct conversion flat-panel detectors using amorphous selenium

Med Phys. 2005 Feb;32(2):488-500. doi: 10.1118/1.1843353.

Abstract

Direct flat-panel detectors using amorphous selenium (a-Se) x-ray photoconductors are gaining wide-spread clinical use. The goal of our investigation is to understand the physical mechanisms responsible for ghosting, i.e., x-ray induced change in sensitivity that results in image persistence, so that the knowledge can be used to consistently minimize ghosting artifacts in a-Se flat-panel detectors. In this paper we will discuss the effect on x-ray sensitivity of charge trapping in a-Se, which is the dominant source for ghosting in a-Se flat-panel detectors. Our approach is to correlate ghosting in electroded a-Se detectors with the trapped charge concentration measured by the "time-of-flight" (TOF) method. All measurements were performed as a function of radiation exposure X of up to approximately 20 R at electric field strength's of E(Se)=5 and 10 V/microm. The results showed that the x-ray sensitivity decreased as a function of X and the amount of ghosting decreased with increasing E(Se). The shape of the TOF curves changed as a result of irradiation in a manner indicating trapped electrons in the bulk of a-Se. The density of trapped electrons n(t) increases as a function of X. A method was developed to determine the values of n(t) in the bulk of a-Se from the TOF measurements, and to predict the corresponding change in x-ray sensitivity. Our results showed that a recombination coefficient consistent with that predicted by Langevin produced good agreement between calculated and measured x-ray sensitivity changes. Thus it can be concluded that the trapping of electrons in the bulk of a-Se and their subsequent recombination with x-ray generated free holes is the dominant mechanism for ghosting in a-Se.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.
  • Validation Study

MeSH terms

  • Artifacts*
  • Computer Simulation
  • Equipment Design
  • Equipment Failure Analysis
  • Models, Theoretical*
  • Radiation Dosage
  • Radiographic Image Enhancement / instrumentation*
  • Radiographic Image Enhancement / methods
  • Radiometry / instrumentation*
  • Radiometry / methods
  • Selenium / chemistry*
  • Selenium / radiation effects*
  • Static Electricity
  • Transducers
  • X-Ray Intensifying Screens*

Substances

  • Selenium