[Digital mammography with high-resolution storage plates (CR) versus full-field digital mammography (CCD) (DR) for microcalcifications and focal lesions -- a retrospective clinical histologic analysis (n = 102)]

Rofo. 2005 Jan;177(1):67-71. doi: 10.1055/s-2004-813722.
[Article in German]

Abstract

Purpose: To determine the diagnostic accuracy of microcalcifications and focal lesions in a retrospective clinical-histological study using high-resolution digital phosphor storage plates (hard copy) and full-field digital mammography (hard copy).

Materials and methods: From May 2003 to September 2003, 102 patients underwent digital storage plate mammography (CR), using a mammography unit (Mammomat 3000 N, Siemens) in combination with a high resolution (9 lp/mm) digital storage phosphor plate system (pixel size 50 microm) (Fuji/Siemens). After diagnosis and preoperative wire localization, full-field digital mammography (CCD) (DR) was performed with the same exposure parameters. The full-field digital mammography used a CCD-detector (SenoScan) (Fisher Imaging) with a resolution of 10 Ip/mm and a pixel size of 50 microm. Five investigators determined the diagnosis (BI-RADS I - V) retrospectively after the operation from randomly distributed mediolateral views (hard copy reading). These results were correlated with the final histology.

Results: The diagnostic accuracy of digital storage plate mammography (CR) and full-field digital mammography (CCD) (DR) was 73 % and 71 % for all findings (n = 102), 73 % and 71 % for microcalcifications (n = 51), and 72 % and 70 % for focal lesions (n = 51). The overall results showed no difference.

Conclusion: Our findings indicate the equivalence of high-resolution digital phosphor storage plate mammography (CR) and full-field digital mammography (CCD) (DR).

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Breast / pathology
  • Breast Diseases / diagnostic imaging*
  • Breast Diseases / pathology
  • Breast Neoplasms / diagnostic imaging*
  • Breast Neoplasms / pathology
  • Breast Neoplasms / surgery
  • Calcinosis / diagnostic imaging*
  • Calcinosis / pathology
  • Diagnosis, Differential
  • Female
  • Follow-Up Studies
  • Humans
  • Mammography / methods*
  • Middle Aged
  • Prospective Studies
  • Radiographic Image Enhancement*
  • Retrospective Studies
  • Sensitivity and Specificity
  • Time Factors