An advanced EPR stopped-flow apparatus based on a dielectric ring resonator

J Magn Reson. 2005 Feb;172(2):312-23. doi: 10.1016/j.jmr.2004.10.015.

Abstract

A novel EPR stopped-flow accessory is described which allows time-dependent cw-EPR measurements of rate constants of reactions involving paramagnetic species after rapid mixing of two liquid reagents. The EPR stopped-flow design represents a state-of-the-art, computer controlled fluid driving system, a miniresonant EPR structure with an integrated small ball mixer, and a stopping valve. The X-band EPR detection system is an improved version of that reported by Sienkiewicz et al. [Rev. Sci. Instr. 65 (1994) 68], and utilizes a resonator with two stacked ceramic dielectric rings separated by a variable spacer. The resonator with the mode TE(H)011 is tailored particularly for conditions of fast flowing and rapidly stopped aqueous solutions, and for a high time resolution. The short distance between the ball mixer and the small EPR active volume (1.8 microl) yields a measured dead time of 330 micros. A compact assembly of all parts results in minimization of disturbing microphonics. The computer controlled driving system from BioLogic with two independent stepping motors was optimized for EPR stopped-flow with a hard-stop valve. Performance tests on the EPR spectrometer ESP 300E from BRUKER using redox reactions of nitroxide radicals revealed the EPR stopped-flow accessory as an advanced, versatile, and reliable instrument with high reproducibility.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electron Spin Resonance Spectroscopy / instrumentation*
  • Equipment Design
  • Rheology / instrumentation*