Low contrast visual acuity in pseudophakic patients implanted with an anterior surface modified prolate intraocular lens

Acta Ophthalmol Scand. 2004 Dec;82(6):718-22. doi: 10.1111/j.1600-0420.2004.00355.x.

Abstract

Purpose: To determine whether implantation of a new intraocular lens (IOL) with a modified prolate anterior surface, designed to reduce the positive spherical aberration of the pseudophakic eyes, results in improved contrast sensitivity assessed in terms of low contrast visual acuity (VA).

Methods: We carried out an intraindividual study of 12 patients with bilateral cataracts, randomized to receive a prolate anterior surface IOL (Tecnis Z9000, Pharmacia) in one eye and a biconvex spherical surface IOL (CeeOn 911 A, Pharmacia) in the other. High and low contrast VA was assessed under photopic conditions, before and after pupil dilation.

Results: After mydriasis, the Technis Z9000 provided significantly better low contrast VA at contrast levels < or = 25%. No significant difference was found under normal pupil conditions.

Conclusion: Our results confirm the hypothesis that the spherical aberration of the eye after cataract surgery can be reduced by an anterior prolate surface IOL.

Publication types

  • Clinical Trial
  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Contrast Sensitivity / physiology*
  • Female
  • Humans
  • Lens Implantation, Intraocular*
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Phacoemulsification
  • Prosthesis Design
  • Pseudophakia / physiopathology*
  • Visual Acuity / physiology*