Analysis of aliphatic and aromatic carbonyl compounds in ambient air by LC/MS/MS

Anal Sci. 2004 Oct;20(10):1459-62. doi: 10.2116/analsci.20.1459.

Abstract

A sensitive liquid chromatograph/tandem mass spectrometric technique (LC/MS/MS) was applied to determine aliphatic and aromatic carbonyl compounds in ambient air. Traces of the carbonyl compounds were sampled by passing through a Sep-Pak DNPH-silica cartridge. Their derivatives were thus eluted with acetonitrile, separated by reversed-phase liquid chromatography and determined by quadrupole tandem mass spectrometry in an atmospheric pressure chemical ionization (APCI) mode with multiple reaction monitoring (MRM). The detection limits (DL) of the carbonyl compounds were 0.8 - 15 ng/m3. A number of the carbonyl compounds were detected at n.d.- 14 microg/m3 levels. The precursor ion scanning analysis was applied to identify the unknown compounds.

MeSH terms

  • Air / analysis*
  • Air Pollutants / analysis*
  • Aldehydes / analysis*
  • Chromatography, Liquid / methods
  • Hydrocarbons, Aromatic / analysis*
  • Ketones / analysis*
  • Mass Spectrometry / methods
  • Sensitivity and Specificity

Substances

  • Air Pollutants
  • Aldehydes
  • Hydrocarbons, Aromatic
  • Ketones