High-performance liquid chromatographic method for simultaneous determination of baicalein and baicalein 7-glucuronide in rat plasma

J Pharm Biomed Anal. 2004 Nov 15;36(3):637-41. doi: 10.1016/j.jpba.2004.07.024.

Abstract

A simple HPLC method with ultraviolet detection was developed for the simultaneous determination of baicalein and its major metabolite, baicalein 7-glucuronide (baicalin) in rat plasma. Following solid phase extraction with HLB cartridge, the analytes were separated using a gradient mobile phase consisting of methanol-acetonitrile-phosphate buffer. The flow-rate was set at 1ml/min and the eluent was detected at 320nm. The method is linear over the studied range of 1-10 and 0.05-1microg/ml for baicalein and baicalin, respectively. The intra-day and inter-day variations of the analysis were less than 6.38%, with relative error ranging from -3.6 to 6%. The limit of quantification in plasma was 0.05microg/ml for both baicalein and baicalin. The developed method proves to be an improved and more reliable method because of its capability for determining baicalein and baicalin simultaneously in a single chromatographic run.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Animals
  • Chromatography, High Pressure Liquid / methods
  • Flavanones / blood*
  • Flavanones / chemistry
  • Flavonoids / blood*
  • Flavonoids / chemistry
  • Male
  • Rats
  • Rats, Sprague-Dawley

Substances

  • Flavanones
  • Flavonoids
  • baicalin
  • baicalein