Development of a static headspace gas chromatographic/mass spectrometric method to analyze the level of volatile contaminants biodegradation

J Chromatogr A. 2004 Sep 3;1048(1):67-71. doi: 10.1016/j.chroma.2004.07.021.

Abstract

Volatile compound biodegradation analysis usually requires the time-consuming step of extraction of the analytes from the matrix using organic solvents or costly radioactive-compounds. Thus, it is desirable to have a simple and fast technique to generate a good evaluation of bacterial biodegradation. The goal of this research was to develop a methodology on the basis of static headspace-gas chromatography/mass spectrometry (HS-GC/MS) to evaluate the level of volatile contaminant biodegradation. The effects of the following parameters were studied: temperature and time of equilibration. The biodegradation experiments were carried out with bacteria inoculation in mineral media in presence of volatile hydrocarbon compounds (toluene, p-xylene, nonane and naphthalene). Autoclaved inoculates were used as control and reference sample. The optimal headspace conditions were observed when the vials were heated at 80 degrees C for 20 min, the syringe at 81 degrees C and an injection volume of 0.4 mL was used. This methodology has the advantage of being relative free from matrix effects.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biodegradation, Environmental
  • Gas Chromatography-Mass Spectrometry / methods*
  • Volatilization