Evaluation of mechanical and corrosion biocompatibility of TiTa alloys

J Mater Sci Mater Med. 2001 Apr;12(4):283-92. doi: 10.1023/a:1011210101895.

Abstract

As-received and heat-treated Ti40Ta and Ti50Ta alloys were evaluated to determine their corrosion as well as mechanical performances and compared to Ti6A14V, a common material utilized for orthopedic (surgical) implants. Anodic potentiodynamic tests performed in Plasmalyte showed that all samples, except for the Ti50Ta specimen aged at 400 degrees C for 3 h gave a curve similar to that of Ti6A14V. Optical and TEM microscopy was performed to determine as-received and heat-treated microstructures. As-received materials showed an alpha precipitate in an alpha+beta and martensite matrix. Samples that were aged at 400 degrees C increased in the density and the length of the alpha precipitate. Vickers hardness measurements were performed to get an approximation of the tensile strengths. Aged Ti40Ta and Ti50Ta specimens produced the highest tensile values when compared to the Ti6A14V material, representing a 31% and 56% increase for the 3 h samples and an 18% and 58% increase for the 10 h samples. Of all the materials studied the Ti50Ta specimen aged for 10 h exhibited the best biocompatibility showing excellent corrosion resistance combined with the highest tensile strength (1089 MPa and 58% harder/stronger than Ti6A14V).