Linear optical coherence tomography system with a downconverted fringe pattern

Opt Lett. 2004 Jul 15;29(14):1644-6. doi: 10.1364/ol.29.001644.

Abstract

Linear optical coherence tomography (LOCT) systems are a simple and robust alternative to time-domain optical coherence tomography systems, but a detector with approximately 10(4) pixels is needed for an imaging depth of 2 mm. We present a new system for LOCT with a special mask attached to the image sensor. The mask essentially performs a downconversion of the spatial frequencies by multiplication with a second spatial frequency. This reduces the fringe frequency of the optical coherence tomography signal so that the signal can be sampled with fewer pixels.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Equipment Design
  • Equipment Failure Analysis
  • Feasibility Studies
  • Image Enhancement / instrumentation*
  • Image Enhancement / methods*
  • Refractometry
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted / instrumentation*
  • Tomography, Optical Coherence / instrumentation*
  • Tomography, Optical Coherence / methods*