Design and construction of shoulder recesses into the beam aperture shields for improved patient positioning at the FiR 1 BNCT facility

Appl Radiat Isot. 2004 Nov;61(5):799-803. doi: 10.1016/j.apradiso.2004.05.030.

Abstract

Improvements have been made at the FiR 1 BNCT facility to ease the positioning of the patient with a tumor in the head and neck region into a lateral neutron beam. Shoulder recesses were constructed horizontally on both sides of the beam aperture. When shoulder recesses are not needed, they are filled with neutron attenuating filling blocks. MCNP simulations using an anthropomorphic human model BOMAB phantom showed that the main contribution to the increase in the effective dose to the patient's body due to the shoulder recesses was from the neutron dose of the arm. In a position when one arm is inside the shoulder recess, the maximal effective dose of the patient was estimated to be 0.7Sv/h. Dose measurements using the twin ionization chamber technique showed that the neutron dose increased on the sides as predicted by the MCNP model but there was no noticeable change in the gamma doses. When making the recesses into the lithium containing neutron shield material tritium contamination was confined using an underpressurized glove box and machine tools with local exhaust. The shoulder recesses give space for more flexible patient positioning and can be considered as a significant improvement of the Finnish BNCT facility.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Boron Neutron Capture Therapy / instrumentation*
  • Equipment Design
  • Finland
  • Head and Neck Neoplasms / radiotherapy
  • Humans
  • Phantoms, Imaging
  • Posture
  • Radiotherapy Planning, Computer-Assisted / instrumentation