Multicenter evaluation of a whole-blood filter that saves platelets

Transfusion. 2004 Aug;44(8):1197-203. doi: 10.1111/j.1537-2995.2004.03350.x.

Abstract

Background: Whole-blood (WB) leukoreduction filters in current use retain the majority of PLTs. A new whole-blood filter, which retains significantly fewer of the PLTs (or saves PLTs [WB-SP]), has been developed. The performance characteristics of the WB-SP filter have been evaluated in a multicenter study.

Study design and methods: A total of 617 units of WB was collected into quadruple bag sets with an integrated WB-SP filter, leukoreduced, and processed into leukoreduced RBCs (LR-RBC), plasma (LR-PL), and buffy coats (LR-BC) from which, pooled, leukoreduced, PLT concentrates (LR-PCs) were produced. Recovery, yield, and residual WBCs were assessed in prepared blood components.

Results: The median residual WBC number in the LR-RBCs was 0.05 x 10(6) (range, <0.05-3.8), exceeding 1 x 10(6) in 0.6 percent of the units. Median Hb content in LR-RBC was 50 g (range, 34-72), reflecting a final RBC recovery of 81 +/- 6 percent. The median WBC content of the LR-PC was 0.05 x 10(6) (range, <0.05-0.28), with none exceeding 1 x 10(6). The median PLT content of the LR-PC, per individual donation, was 6.4 x 10(10) (range, 4.1-10.7), representing a final recovery of 62 +/- 10 percent. The mean FVIII activity was 104 +/- 25 percent and 83 +/- 11 percent in plasma separated from fresh or overnight stored WB, respectively.

Conclusion: Use of the WB-SP filter makes it possible to obtain three leukoreduced blood components with only one filtration step. The WB-SP filter showed good leukoreduction performance and recovery of all blood components including PLTs.

Publication types

  • Multicenter Study

MeSH terms

  • Blood Component Removal / instrumentation*
  • Blood Platelets*
  • Filtration / methods*
  • Humans
  • Leukocytes*