Determination of rare earth elements in geological samples by inductively coupled plasma atomic emission spectrometry with flow injection liquid-liquid extraction

Anal Sci. 2003 Dec;19(12):1625-9. doi: 10.2116/analsci.19.1625.

Abstract

A direct sampling with organic solvent extracts for simultaneous multi-element determination implemented with inductively coupled plasma atomic emission spectrometry (ICP-AES) associated with a flow injection liquid-liquid extraction (FI-LLE) sample preconcentration method was studied. The "robustness" of the plasma discharge with tributyl phosphate (TBP) loading was diagnosed by using the Mg II 279.55 nm and Mg I 285.21 nm lines intensity ratio. A FI-LLE preconcentration system for rare earth elements (REEs)-nitrate-TBP was established by using a laboratory-designed phase separator. For these elements, an average sensitivity enhancement factor of 64 was obtained with respect to ICP-AES sampling with aqueous solutions. The precision of the method was characterized by a relative standard deviation (%RSD) of 1.8 - 5.2%. A throughput of 27 samples per hour can be achieved with an organic solvent consumption of less than 200 microl per determination. Good results were obtained for the analysis of standard reference materials.