Preliminary evaluation of image reconstruction by ordered-subset expectation maximization in thallium-201 lung scintigraphy

Ann Nucl Med. 2003 Oct;17(7):619-24. doi: 10.1007/BF03006680.

Abstract

Objectives: We observed whether clearer tumor delineation and greater tumor to non-tumor (T/N) count ratios could be obtained using an iterative ordered-subsets expectation maximization (OSEM) algorithm than conventional filtered-back projection algorithm (FBP) in the image reconstruction of thallium-201 (201Tl) lung scintigraphy.

Methods: In 29 patients with lung cancer and phantom studies, tomograms were reconstructed using FBP and OSEM algorithms, with and without a prefilter (Butterworth filter: BW), whose cut-off frequencies were 0.10 cycles/pixel for FBP and 0.10 and 0.17 cycles/pixel for OSEM. Visual interpretation and tumor to non-tumor (T/N) count ratios were obtained and compared.

Results: Without a prefilter, T/N ratios from OSEM and FBP were 1.89 +/- 0.31 (early) and 2.00 +/- 0.54 (late) for OSEM, 1.90 +/- 0.33 (early) and 2.05 +/- 0.59 (late) for FBP, respectively. The OSEM reconstruction without prefiltering showed clearer tumor contours than FBP without a prefilter. Incorporation of BW showed visually low-noised images but decreased T/N ratios in both reconstructions with BW (0.10 cycles/pixel). No greater T/N ratios were obtained by OSEM than FBP, with or without prefiltering. With BW with a cut-off frequency of 0.17 cycles/pixel, the same T/N ratios as those without BW were obtained. The tumor model sized 0.9 cm in the phantom study was invisible in both OSEM and FBP reconstructions without a prefilter, but visible with a prefilter. The influence of prefiltering on T/N ratios was also observed in phantom studies.

Conclusions: Visually improved tumor delineation could be obtained in OSEM reconstruction without a prefilter as compared to FBP reconstruction without a prefilter for tumors greater than 2 cm. Prefiltering should be incorporated into OSEM reconstruction in diagnosing small tumors. However, the influence of prefilter (BW) setting on semi-quantitative interpretation needs further discussion.

Publication types

  • Clinical Trial
  • Comparative Study
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Feasibility Studies
  • Female
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Lung Neoplasms / diagnostic imaging
  • Lung Neoplasms / pathology
  • Male
  • Middle Aged
  • Neoplasm Staging / methods
  • Pilot Projects
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Solitary Pulmonary Nodule / diagnostic imaging*
  • Solitary Pulmonary Nodule / pathology
  • Thallium*
  • Tomography, Emission-Computed, Single-Photon / methods*

Substances

  • Radiopharmaceuticals
  • thallium chloride
  • Thallium