Capillary electrophoresis coupled to mass spectrometry to establish polypeptide patterns in dialysis fluids

J Chromatogr A. 2003 Sep 26;1013(1-2):157-71. doi: 10.1016/s0021-9673(03)00712-x.

Abstract

Combination of capillary electrophoresis with mass spectrometry (CE-MS) allows generation of polypeptide patterns of body fluids. In a single CE-MS (45 min) run more than 600 polypeptides were analyzed in hemodialysis fluids obtained with different membranes (high-flux/low-flux). Larger polypeptides (M(r) > 10 000) were almost exclusively present in high-flux dialysates only, while in low-flux dialysates additional small polypeptides were detected. Comparison to the normal urine pattern yielded a surprisingly low consensus: a number of polypeptides present in urine were missing. We established a fast and sensitive technique, easily applicable to the monitoring of different modalities of dialyzers.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Dialysis
  • Electrophoresis, Capillary / methods*
  • Peptides / analysis*
  • Spectrometry, Mass, Electrospray Ionization / methods*

Substances

  • Peptides