Measurement of ultrasonic power using an acoustically absorbing well

Ultrasound Med Biol. 2003 Oct;29(10):1507-13. doi: 10.1016/s0301-5629(03)01036-6.

Abstract

This paper describes a quick and cost-effective method for constructing a radiation force balance for measuring ultrasonic output power. It utilises a target manufactured from a high-quality acoustical absorber material. The target geometry is in the form of a cup or well that is water-filled and placed directly on the pan of a top-loading chemical balance, thus overcoming the need for the traditional gantry arrangement found in the majority of commercially available balances. The face of the transducer is placed directly in the water contained within the well. This simplification reduces time spent in setting up a balance for measurement, and targets can be manufactured to any required geometry and used on any suitable top-loading balance to measure output power. Within this study, the performance of the absorbing well method was evaluated over the frequency range of 1 MHz to 5 MHz, for acoustic power levels up to 1 W. Power measurements on three transducers were compared with measurements made on the National Physical Laboratory (NPL) primary standard radiation force balance and good agreement is demonstrated between the two systems. At a power of 50 mW, using a chemical balance of resolution 0.1 mg, typical type A (random) uncertainties were +/- 2.0% when expressed at the 95% confidence level.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorption
  • Acoustics / instrumentation
  • Equipment Design
  • Humans
  • Quality Control
  • Reproducibility of Results
  • Transducers
  • Ultrasonic Therapy / instrumentation
  • Ultrasonic Therapy / standards*
  • Ultrasonography / instrumentation
  • Ultrasonography / standards*