Obtaining location of I-125 seed sources for the prostate brachytherapy

Igaku Butsuri. 2003;23(2):132-9.

Abstract

Treatment planning for ultrasound-guided transperitoneal I-125 permanent prostatic implants is a time consuming task, because this treatment uses many seeds (e.g. 50-80) and there are many solutions (seed positions) for optimizing the dose distribution. In conventional treatment planning for brachytherapy the locations of sources are usually input into a computer manually with reference to two X-ray films. We use so many sources in this treatment that we can hardly obtain accurate location of each source. That is, the identification of each seed with two X-ray films is very difficult. Even though we can use X-ray CT images for deciding on the source position, there are artifacts due to the high attenuation material. This paper proposes a method for obtaining accurate locations of radioactive sources from some projection data (e.g. RPO 60, 40, 20, LPO 60, 40, 20 and 0 deg.) with a reconstruction method. In our method for obtaining actual source positions we use an iterative image reconstruction technique. The results obtained by experiments with a simple prostate phantom and the RANDO phantom showed that the locations of dummy seeds were accurately decided with a reconstructed image by six or seven projection data.

MeSH terms

  • Brachytherapy*
  • Humans
  • Male
  • Phantoms, Imaging
  • Prostatic Neoplasms*
  • Seeds
  • Tomography, X-Ray Computed