Triphasic waveforms are superior to biphasic waveforms for transthoracic defibrillation: experimental studies

J Am Coll Cardiol. 2003 Aug 6;42(3):568-75. doi: 10.1016/s0735-1097(03)00656-9.

Abstract

Objectives: Our objective was to evaluate the efficacy of triphasic waveforms for transthoracic defibrillation in a swine model.

Background: Triphasic shocks have been found to cause less post-shock dysfunction than biphasic shocks in chick embryo studies.

Methods: After 30 s of electrically induced ventricular fibrillation (VF), each pig in part I (n = 32) received truncated exponential biphasic (7.2/7.2 ms) and triphasic (4.8/4.8/4.8 ms) transthoracic shocks. Each pig in part II (n = 14) received biphasic (5/5 ms) and triphasic shocks (5/5/5 ms). Three selected energy levels (50, 100, and 150 J) were tested for parts I and II. Pigs in part III (n = 13) received biphasic (5/5 ms) and triphasic (5/5/5 ms) shocks at a higher energy (200 and 300 J). Although the individual pulse durations of these shocks were equal, the energy of each pulse varied. Nine pigs in part I also received shocks where each individual pulse contained equal energy but was of a different duration (biphasic 3.3/11.1 ms; triphasic 2.0/3.2/9.2 ms).

Results: Triphasic shocks of equal duration pulses achieved higher success than biphasic shocks at delivered low energies: <40 J: 38 +/- 5% triphasic vs. 19 +/- 4% biphasic (p < 0.01); 40 to <50 J: 66 +/- 7% vs. 42 +/- 7% (p < 0.01); and 50 to <65 J: 78 +/- 4% vs. 54 +/- 5% (p < 0.05). Shocks of equal energy but different duration pulses achieved relatively poor success for both triphasic and biphasic waveforms. Shock-induced ventricular tachycardia (VT) and asystole occurred less often after triphasic shocks.

Conclusions: Triphasic transthoracic shocks composed of equal duration pulses were superior to biphasic shocks for VF termination at low energies and caused less VT and asystole.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Electric Countershock / methods*
  • Electrophysiology
  • Models, Animal
  • Swine
  • Treatment Outcome
  • Ventricular Fibrillation / therapy*