Development of electrochemical methods for determination of tramadol--analytical application to pharmaceutical dosage forms

J Pharm Biomed Anal. 2003 Aug 8;32(4-5):975-81. doi: 10.1016/s0731-7085(03)00199-7.

Abstract

A square-wave voltammetric (SWV) method and a flow injection analysis system with amperometric detection were developed for the determination of tramadol hydrochloride. The SWV method enables the determination of tramadol over the concentration range of 15-75 microM with a detection limit of 2.2 microM. Tramadol could be determined in concentrations between 9 and 50 microM at a sampling rate of 90 h(-1), with a detection limit of 1.7 microM using the flow injection system. The electrochemical methods developed were successfully applied to the determination of tramadol in pharmaceutical dosage forms, without any pre-treatment of the samples. Recovery trials were performed to assess the accuracy of the results; the values were between 97 and 102% for both methods.

MeSH terms

  • Dosage Forms
  • Electrochemistry
  • Technology, Pharmaceutical / methods*
  • Tramadol / administration & dosage
  • Tramadol / analysis*
  • Tramadol / chemistry*

Substances

  • Dosage Forms
  • Tramadol