Comparison of the haemodynamic effects of right ventricular outflow-tract pacing with right ventricular apex pacing: a quantitative review

Europace. 2003 Jul;5(3):275-8. doi: 10.1016/s1099-5129(03)00031-x.

Abstract

The right ventricular apex has been used for cardiac stimulation because this position is easily accessible and is associated with a stable position of the electrode with a low dislodgement rate. This position, however, is associated with a dyssynchronous left ventricular contraction with subsequent deleterious haemodynamic effects. Alternative stimulation sites have been studied extensively because of a potentially better haemodynamic effect compared with right ventricular apex pacing. Using a Cochrane search strategy, nine studies were selected to analyze the haemodynamic effects of right ventricular outflow-tract pacing. The results of these studies (n=217) were pooled and indicated a significantly better haemodynamic effect (odds ratio 0.34, confidence interval 0.15-0.53) compared with right ventricular apex pacing. Therefore, these data suggest that right ventricular outflow-tract pacing may offer a modest but significant benefit over right ventricular apex pacing in patients selected for pacemaker implantation on the basis of symptomatic bradyarrhythmias.

Publication types

  • Comparative Study
  • Review

MeSH terms

  • Bradycardia / physiopathology
  • Bradycardia / therapy*
  • Cardiac Pacing, Artificial / adverse effects*
  • Heart Ventricles / physiopathology*
  • Hemodynamics / physiology*
  • Humans
  • Ventricular Dysfunction, Left / etiology*
  • Ventricular Dysfunction, Left / physiopathology*