Experimental investigations of image quality in X-ray mammography with a conventional screen film system (SFS) and a new full-field digital mammography unit (DR) with a-Se-detector

Rofo. 2003 Jun;175(6):766-8. doi: 10.1055/s-2003-39936.

Abstract

Purpose: Determination of image quality of a new digital mammography system with an a-Se-detector.

Material and method: Radiograms of the Wisconsin Mammographic Random Phantom, Modell 152 A (Radiation Measurements Inc., Wisconsin) were acquired using a conventional film screen and a digital system.

Results: With the conventional film screen system there were detected 191/a 38.2 and with the digital mammography 219/a 43.8 of 225 possible details.

Conclusion: Based on these results there is the possibility to replace the conventional film screen system by the digital mammography with a-Se-detector--this has to be confirmed in further clinical trials.