Rapid fluorimetric assay for primary amine groups in water samples

Anal Bioanal Chem. 2003 Jul;376(6):918-22. doi: 10.1007/s00216-003-2004-0. Epub 2003 Jun 17.

Abstract

Bond Elut C(18) solid-phase extraction cartridges were used for pre-concentration followed by derivatization with o-phthaldialdehyde-N-acetylcysteine (OPA-NAC) of primary amines in water. Optimal conditions were: conditioning the cartridges with borate buffer pH 10.4, retention of the primary amines, addition of the OPA-NAC(3.7 mmol L(-1)) 1:1 molar ratio and borate buffer pH 8, elution of the isoindol with MeOH-borate buffer (9:1) pH 10.2 and fluorescence measurement. The equations of the calibration graphs for methylamine, ethylamine, propylamine, butylamine, pentylamine, and beta-phenylethylamine at lambda(excitation)=330 nm and lambda(emission)=440 nm, in the optimal conditions are presented. The solid-phase extraction procedure improved ten times the detection limits of the solution derivatization. Those values are in the 0.01-0.06 mg L(-1) interval in function of the amine. Also, it is possible to estimate the total primary aliphatic amine concentration in water, expressed as molar concentration of -NH(2) group or -NH(2)-N mg L(-1). On the basis of these studies, the method was applied for the determination of primary amino groups in tap, ground, factory and source water samples.