In-use evaluation of Perasafe compared with Cidex in fibreoptic bronchoscope disinfection

J Hosp Infect. 2003 May;54(1):46-51. doi: 10.1016/s0195-6701(03)00072-0.

Abstract

The mycobactericidal activity of Perasafe (0.26% peracetic acid) was compared with that of Cidex (2% alkaline glutaraldehyde) by an in-use test. Fibreoptic bronchoscopes were artificially contaminated with Mycobacterium tuberculosis or Mycobacterium avium-intracellulare in sputum and, after manual pre-cleaning with a neutral soap, 10 and 20 min disinfection periods were tested. Perasafe was as effective as Cidex, thus requiring a 10 min disinfection period against M. tuberculosis and 20 min against M. avium-intracellulare. The results demonstrate that Perasafe is an effective disinfectant for use in reprocessing fibreoptic bronchoscopes.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bronchoscopes / microbiology*
  • Cross Infection / prevention & control
  • Disinfectants / administration & dosage
  • Disinfectants / pharmacology*
  • Equipment Contamination / prevention & control
  • Glutaral / administration & dosage
  • Glutaral / pharmacology*
  • Humans
  • Infection Control / methods
  • Mycobacterium avium Complex / drug effects*
  • Mycobacterium tuberculosis / drug effects*
  • Peracetic Acid / administration & dosage
  • Peracetic Acid / pharmacology*
  • Sputum / microbiology

Substances

  • Disinfectants
  • Peracetic Acid
  • Glutaral